Measurementof Serum AlbuminbyThree Different Methods
Taghreed,AbdulHameedDepartment of Biochemistry,College of Medicine,University of Tikrit,Tikrit,Iraq
ABSTRACT
In this study ,serum albumin level was determined in 61 blood samples using three different methods ,cellulose acetate electrophoresis (CAE), salt fractionation and bromocresolgreen(BCG) . The values of serum albumin (mean (95% confidence interval)) measured in gm/L by the above methods were 38.95 (1.64) , 38.13 (1.76) and 38.11 (1.8) respectively .There were no significant differences in the level of serum albumin determined by these methods (P≥0.05). Also, these methods showed statistically significant differences in the level of serum albumin determined by these methods (P≤ 0.001) ,As the BCG method give a relatively inaccurate albumin levels in abnormal sera therefore ,it is recommended that the BCG is to be used only as a screening method for serum albumin measurement for its simplicity and rapidity
قياس مستوى الالبومين في الدم با ستخدام ثلاث طرق مختلفة
تغريد عبد الحميد ,كلية الطب, فرع الكيمياء الحياتية , جامعة تكريت, تكريت العراق
الخلاصة
في هذه الدراسة ,تم قياسمستوى الالبومين في الدم في 61 نموذج با ستخدام ثلاث طرق مختلفة الترحيل الكهربائي , الترسيب التجزيئي , وطريقة البروموكريسول كرين (BCG),وكانت قيمالالبومين]( 95%CL) X–[مقاسا ب غم/لترللطرق الثلاثة هي38.95(1.64),38.13(1.76))38.11(1.8) على التوالي .
لم يكن هناك ةاختلاف احصائي مهم في مستوى الالبومين مقاسا بالطرق الثلاثة (0.05p≥). أظهرت الدراسة وجود ترابط موجب مهم احصائيا بين الطرق الثلاثة( 0.001P ≤) ونتيجة لكون طريقة(BCG) تعطي نتائج البومين غير دقيقة نسبيا في النماذج غيرالطبيعية, لذلك نوصي باستعمالها فقط كطريقة مسح لقياس مستوى الالبومين وذلك لبساطتها وسرعتها.