قياس مستوى الالبومين في الدم  با ستخدام ثلاث طرق مختلفة

Measurementof Serum AlbuminbyThree Different Methods

Taghreed,AbdulHameedDepartment of Biochemistry,College of Medicine,University of Tikrit,Tikrit,Iraq

ABSTRACT

In this study ,serum albumin level was determined in  61  blood  samples using three different methods ,cellulose  acetate electrophoresis (CAE), salt  fractionation and  bromocresolgreen(BCG) . The values of  serum  albumin  (mean (95% confidence interval)) measured  in gm/L by the above methods were 38.95 (1.64) , 38.13 (1.76) and 38.11 (1.8) respectively .There were  no significant  differences in the level  of  serum  albumin  determined by these methods (P≥0.05). Also, these methods  showed  statistically  significant  differences in the level of serum albumin determined  by these methods  (P≤ 0.001) ,As the BCG method  give a relatively  inaccurate albumin levels in abnormal sera  therefore ,it is recommended that  the BCG is to be used only  as a screening method for  serum albumin measurement for its simplicity and rapidity

قياس مستوى الالبومين في الدم  با ستخدام ثلاث طرق مختلفة

تغريد عبد الحميد ,كلية الطب, فرع الكيمياء الحياتية , جامعة تكريت,  تكريت العراق

الخلاصة

في هذه الدراسة ,تم قياسمستوى الالبومين في الدم في 61  نموذج با ستخدام ثلاث طرق مختلفة  الترحيل الكهربائي , الترسيب التجزيئي , وطريقة البروموكريسول كرين (BCG),وكانت قيمالالبومين]( 95%CLX[مقاسا ب غم/لترللطرق الثلاثة هي38.95(1.64),38.13(1.76))38.11(1.8) على التوالي .

لم يكن هناك ةاختلاف احصائي مهم في مستوى الالبومين مقاسا بالطرق الثلاثة (0.05p≥). أظهرت الدراسة وجود ترابط موجب مهم احصائيا بين الطرق الثلاثة( 0.001P ≤) ونتيجة لكون  طريقة(BCG) تعطي نتائج البومين غير دقيقة نسبيا في النماذج غيرالطبيعية, لذلك نوصي باستعمالها  فقط كطريقة مسح لقياس مستوى الالبومين وذلك لبساطتها وسرعتها.

تحميل البحث